High Resistance Meter
HR2 - Alpha Lab, Inc. (Trifield)
The Model HR2 meter measures resistance in nine ranges from 1 ohm to 1.9999 Tera ohms (1.9999x1012 ohms), and conductance in a single range of 0.00 to 199.99 pS (pico siemens, or pico mhos), all at low voltage of < 2 volts across the sample. Basic accuracy is +/- 2% of reading +/- 1 count. For resistance higher than 1.9999 Tera ohms, the conductance range can be used to measure conductance down to 0.01 pS (100 T ohms). The meter has a conductance range because samples with simultaneously high resistance and high capacitance settle faster to their final value on conductance (usually under one second), than on resistance (several seconds), and conductance is also capable of detecting higher resistance than the resistance ranges can. Note that resistance in units of T ohms is the inverse of conductance in pS, so that the resistance can be calculated from the conductance. For example, 100 pS is 0.01 T ohm (10 G ohm), and the highest readable conductance of 199.99 pS is essentially 5 G ohms.