Dual Vapor Gravimetric Sorption Analyzer
DVS Resolution - Surface Measurement Systems
DVS Resolution’s key feature is the combined ability to precisely control and measure temperature and relative humidity while recording the highest resolution changes in mass. DVS Resolution uses a dry carrier gas typically nitrogen or compressed air for its operation. Digital mass flow controllers regulate flows of dry and saturated gases. Relative humidity is generated by precisely mixing dry and saturated gas flows in desired flow ratios which produce expected relative humidity. In a typical experiment a known concentration of water vapor is delivered over a sample placed inside the sample pan connected to the Surface Measurement Systems Ultrabalance™ measuring realtime mass changes caused by sorption or desorption of water molecules. Prior to sorption measurements, the sample can be in-situ preheated/dried at temperatures of up to 300oC.